Introducción
● DO measurement has ATC, automatic salinity compensation and automatic air pressure compensation. |
●Equips new type of DO electrode which has function of automatic temperature compensation and automatic salinity compensation as well as automatic air pressure compensation inside meter to make DO measuring mode has temperature, air pressure and salinity automatic compensation, easy to use, accurate to measure. |
●Meter meets with the requirement of international GLP standards. Its features up to automatic calibration, automatic temperature compensation, data storage, timing measurement, clock display, RS-232 output, function setting and self-diagnose etc. |
●There are three different measurement modes: stable display mode; time measurement mode; continuous measurement mode. |
●Electrode adopts combination diaphragm cap, assembling sensitive membrane with diaphragm cap, which very convenient to replace the sensitive membrane. The user only needs to screw the new membrane cap on the electrode. |
● Meter has the only product serial number. |
● Dustproof and waterproof meter meets IP54 standards. All sockets are seal protected by the silica gel caps. |
Especificaciones
Dissolved Oxygen |
Measuring range
Resolution
Accuracy
Response time
Residual current
Temp. compensation range
Salinity compensation range
Air pressure compensation
Automatic calibration
Electrode type |
(0 ~ 40.00)mg/L(ppm) (0 ~ 200.0)%
0.1/0.01 mg/L(ppm) 1/0.1 %
Meter:±0.10 mg/L,connect meter with electrode:±0.40 mg/L
≤30 s(25℃, 90% response)
≤0.1 mg/L
(0 ~ 45)℃(automatic)
(0 ~ 45)ppt(automatic)
(80 ~ 105)kPa(automatic)
Air saturated by water; water saturated by air
Polarogram type |
Temperature |
Measuring range
Resolution
Accuracy |
-10℃~ 110℃
0.1℃
5~ 60℃range:±0.4℃ other range:±0.8℃ |
Other Technical Parameters: |
Data storage
Storage content
Power
Communication connector
Size and weight
Quality and safety certification |
300 groups
Measuring value series number, measuring value, temperature, ATC or MTC state, measuring date and time
DC9V/300mA
RS232
160 × 190 × 70mm/880g
ISO9001:2008, CE and CMC |
Working Condition |
Environment temperature
Environmental humidity
IP rating |
5 ~ 35 ℃(0.01grade) 15 ~ 30 ℃(0.001grade)
≤75%
IP54 Dustproof and waterproof |
El kit estándar incluye
1. Model MP516 DO meter | 1unit |
2. Model 602 flexible electrode holder | 1pc |
3. Model 901 intelligent stirrer | 1unit |
4. Model 500 DO electrode | 1pc |
5. DO502 DO electrode inner solution (30mL) | 1bottle |
6. DO503 membrane cap (for DO electrode use) | 3pcs |
7. Cathode polishing paper | 2 small pieces |
8. 9V power adapter (for meter use) | 1pc |
9. 6V power adapter (for stirrer use) | 1pc |
10. RS-232 communication cable | 1pc |
11. MP516 communication software CD | 1pc |
12. B628 stirring bead | 2pcs |
13.AA battery | 4 units |
14. Operation manual | 1unit |
15. Brief operation instruction | 1 unit |